pylabrobot.azenta.fluidx.intellixcap96.FluidXIntelliXcap96.eject_caps#

async FluidXIntelliXcap96.eject_caps(timeout: float = 30.0) None#

Home the cap head, dropping any caps held on the cap drivers.

Closes the tray first if it is open, and opens the safety door. Only available in manual recovery mode, and the instrument stays there afterwards. The caps fall wherever the head is standing, so clear the deck and read waste() before using this to get rid of caps deliberately.

Parameters:

timeout (float)

Return type: