pylabrobot.gyger.CertusValve.dispense#
- async CertusValve.dispense(volume: float, *, correction_factor: float | None = None) None#
Dispense a calibrated volume in µL at the current position.
Attach a
ValveCalibrationtocalibrationfirst. The pressure setpoint and actual pressure must match that calibration within 0.01 bar. Supply the channel’s CERTUS CONTROL volume correction factor when applicable; omitting it uses the configured channel factor (initially 1.0). The factor is applied before interpolation.