pylabrobot.mettler_toledo.MTSICSDriver.request_net_weight_with_status#

async MTSICSDriver.request_net_weight_with_status() MettlerToledoResponse#

Query net weight with unit and weighing status in one call. (SIS command)

Response data fields:

  • data[0] = State: 0=stable, 1=dynamic, 2=stable inaccurate (MinWeigh), 3=dynamic inaccurate, 4=overload, 5=underload, 6=error

  • data[1] = Net weight value

  • data[2] = Unit code: 0=g, 1=kg, 3=mg, 4=ug, 5=ct, 7=lb, 8=oz, etc.

  • data[3] = Readability (number of decimal places, 0-6)

  • data[4] = Step: 1, 2, 5, 10, 20, 50, or 100

  • data[5] = Approval: 0=standard (not approved), 1=e=d, 10=e=10d, 100=e=100d, -1=unapproved

  • data[6] = Info: 0=without tare, 1=net with weighed tare, 2=net with stored tare

Return type: