pylabrobot.thermo_fisher.kingfisher.flex.KingFisherFlex#
- class pylabrobot.thermo_fisher.kingfisher.flex.KingFisherFlex(serial_number: str | None = None, timeout: float = 30.0, operation_timeout: float = 120.0, *, name: str = 'kingfisher_flex')#
Bases:
objectRun supported 96DW BindIt exports over USB HID.
Hardware validation covers firmware 1.01.0 on macOS, including mixing, bead collection/release, drying, 37 C heating, loops and host confirmations. Firmware completion does not verify correct tip-comb attachment.
setupidentifies the device;initializemoves its axes. Commands execute serially.stopdisconnects;abortrequests a motion stop. Execution uses local BDZ exports; onboard protocol management is unsupported.- Parameters:
serial_number (str | None) – Select one instrument when multiple Flex devices are attached.
timeout (float) – HID write/acknowledgment and abort deadline, in seconds.
operation_timeout (float) – Base motion deadline and margin for timed operations, in seconds.
name (str) – Resource-name prefix. Use distinct names for instruments in the same resource tree.
Attributes
Whether the caller has reconciled the current tree with physical contents.
Return the persistent holder accessible at the single loading station.
Return the printed tray at the loading station, or None during uncertain motion.
Return the persistent holder at processing, without moving its assigned plate.
Return the last completed MOR target, or None when physical state is unknown.
Return the session transcript, including messages after interruption.
Methods
__init__([serial_number, timeout, ...])Configure a Flex without enumerating or opening USB devices.
abort()Request PC abort without waiting for the active command or operator pause.
events()Consume session events from one task until disconnect or failure.
Initialize axes and disable heating; this can move axes and present tray 4 for loading.
Confirm that every holder's assignment matches the physical carousel.
run_protocol(protocol, *[, ...])Stream a supported BindIt export and await terminal completion.
setup()Identify the instrument without EAK, fault acknowledgment or motion commands.
stop()Disconnect and invalidate position; never claim that disconnect stops motion.
wait_for_physical_start(action)Send the selected action's original PAU and await its firmware completion once.