pylabrobot.formulatrix.mantis.driver.stage_transform#
The stage transform: where a point on the wellplate is in the arm frame. Pure maths.
The wellplate frame’s origin is the wellplate’s back-left corner, the plate holder’s datum, with x to the right, y to the front and z down from the unit’s reference height above the wellplate’s base, in mm. A point is turned and moved by the arm type, moved by the stage origin, its x and y mapped by the projective transformation matrix and its z tilted; never through a fitted homography.
Functions
|
A point of the wellplate frame in the arm frame, z as Z's position down from its flag. |