pylabrobot.bmg_labtech.vantastar.driver.master.VANTAstarDriver.build_fluorescence_run#
- VANTAstarDriver.build_fluorescence_run(monochromators: Sequence[MonochromatorSettings], *, gains: Sequence[int], focal_height: float, flashes: int, settling_time: float, enhanced_dynamic_range: bool = False, geometry: PlateGeometry = PlateGeometry(length=127.76, width=85.48, first_x=14.38, first_y=11.24, last_x=113.38, last_y=74.24, columns=12, rows=8, well_size=6.58, well_shape=0), well_mask: bytes | None = None, well_scan: Literal['point', 'orbital', 'spiral', 'matrix'] = 'point', scan_diameter: int = 3, corner: Literal['top_left', 'top_right', 'bottom_left', 'bottom_right'] = 'top_left', reading_direction: Literal['column_major', 'row_major'] = 'row_major', bidirectional: bool = True) bytes#
RUN fields for a top fluorescence-intensity endpoint read (MeasurementMode 0).
- Parameters:
monochromators (Sequence[MonochromatorSettings]) – one
MonochromatorSettingsper chromatic.gains (Sequence[int]) – one gain per chromatic; ignored (sent as 0) with enhanced dynamic range.
focal_height (float) – mm, in 0.1 mm steps.
flashes (int) – flashes per well.
settling_time (float) – seconds to settle before reading each well (20 ms steps).
enhanced_dynamic_range (bool) – read with enhanced dynamic range.
geometry (PlateGeometry) – the plate.
well_mask (bytes | None) – the wells to read (
encode_well_mask); None reads every well.well_scan (Literal['point', 'orbital', 'spiral', 'matrix']) – point, orbital or spiral (
encode_well_scan); scan_diameter: its diameter, mm.corner (Literal['top_left', 'top_right', 'bottom_left', 'bottom_right']) – the reading order (
encode_reading_mode).reading_direction (Literal['column_major', 'row_major']) – the reading order (
encode_reading_mode).bidirectional (bool) – the reading order (
encode_reading_mode).scan_diameter (int)
- Return type: